Gerätebeschreibung

System description:
Mixing chamber (injection chamber) to inject a known volume of a component to generate a gas mixture with known concentration. The process is the primary standard to generate gas mixtures to calibrate gas analyzers. The development was for the most part carried out by the National EU Clean Air Reference Lab, a division of the Federal Environmental Agency, and is documented in ISO 6144 as a standard.

The mixing chamber features sensors for the precise determination of the temperature and pressure. The values of the sensors can be directly imported to the laboratory information and management software IC-LIMS-2001 .

Methode der Prüfgaserzeugung

Injection of a known volume of a pure component into a known volume of carrier gas.

Konzentrationseinstellung

By varying the volume of the injected component.

Anmerkungen

Additional data concerning static volumetric injection available upon request.